previous section: 6/7 page: 5/12 next

6Reconstruction of the orbital floor defect

Contoured edges of the plate

… and contoured to achieve the required shape, and accommodate key anatomical structures (nasolacrimal duct, infraorbital nerve, and optic nerve)

It is advisable not to extend the implant further posterior than 1 cm anterior to the optic canal entrance (if the posterior support to the orbit can be reached).