previous section: 8/16 page: 4/5 next

8Choice of implant

Nail diameter

It is important to measure the medullary diameter at the mid portion of the femur, which represents the narrowest segment of the medullary canal (isthmus).

The inner cortical edge should touch the inner numbered disk of a ruler aperture. In the illustration an inner cortical diameter of 14 mm is shown.