7Plate application

Short oblique head fracture of the proximal phalanx – Lag screw with protection plate

Align the plate with the diaphysis

Before inserting distal screw adjacent to the blade, ensure that the plate is in line with the phalangeal diaphysis in the lateral view by rotating it around the long axis of the blade.

Insert distal screw
The distal screw, adjacent to the blade, is then inserted in a neutral position.
The screw should just engage the far cortex.

Note