7Plate application

Short oblique head fracture of the proximal phalanx – Lag screw with protection plate

Drill neutral proximal hole

Use a 1.1 mm drill bit to prepare a neutral thread hole for the first screw at the proximal end of the plate.
To improve stability, drill this hole at a slightly divergent angle.