Reduction

Insert a second K-wire

A second 1.6 mm K-wire is inserted parallel to the first one into the posterior portion of the tip of the lateral malleolus.

These two parallel K-wires stabilize the fracture against rotation and will anchor the figure-of-eight wire distally.

Check the reduction under image intensification.

Note

Care must be taken to avoid intraarticular insertion of the K-wires - check specifically with the image intensifier.