Antegrade nailing (trochanteric)

Antegrade nailing femoral shaft – trochanteric entry point – Guide wire

Determination of entry point and guide-wire insertion

A finger is inserted to palpate the greater trochanter. The 2.8 mm guide wire, inserted under image intensifier control, enters the medullary canal at a slightly oblique angle. Introduction is manual, using the universal chuck with T-handle.

Verify the entry point by both AP and lateral views with image intensifier.

(Image shows external fixator used as temporary fixation device.)