Pediatric proximal femur 31-M/2.1 III

Pediatric hip locking plate

1. Preliminary remarks

Introduction

Children's hip fractures have historically had a bad prognosis. Accurate stable fixation has been one of the factors identified to improve the outcome.

Stable fractures have a better prognosis than displaced fractures.

The main problems have been avascular necrosis (AVN) due to disruption of the blood supply, nonunion and malunion due to unstable fixation.

The objective is timely fixation without disruption of the blood supply.

Angularly stable fixation is the solution to malunion. This is the rationale for using a pediatric hip locking plate.

Implant selection

The choice of implant size depends on the age and weight of the child.

Generally, a 130° angled plate is appropriate for fracture fixation as it corresponds most closely to the child's femoral neck-shaft angle. It offers the possibility of three parallel screws to the neck.

A plate with three shaft holes is sufficient for intracapsular fractures. If the fracture is extracapsular and involves the femoral shaft, pediatric hip locking plates are available with additional shaft holes.

Longer locking screws, not included in the standard pediatric hip locking plate set, may be required.

Hardware placement

During preoperative planning, a decision should be made whether fixation needs to cross the physis. This decision will depend on the level of the fracture.

Note: It is more important to stabilize the fracture than to avoid crossing the physis.

2. Patient preparation and approaches

Patient preparation

This procedure is normally performed with the patient in a supine position .

See also the additional material on preoperative preparation .

Approaches

For this procedure the following approaches may be used:

3. Fixation

Insertion of guide wires

The guide wires for the pediatric hip locking plate are inserted.

This can be done using the dedicated angled guide, or by placing the wires through the plate with the screw-in guides attached to it.

An anteversion wire along the front of the femoral neck can be useful.

A 2.0 mm guide wire should be inserted high in the neck starting at least 5 mm below the trochanteric apophysis. This wire can be inserted using the angled guide, or freehand.

The position should be checked radiologically. On the lateral view, the wire should be in the middle of the femoral neck.

After the initial 2.0 mm wire, two 2.8 mm wires will be inserted using the dedicated angled guide.

The 2.8 mm wires will sit just above the mid-point of the femoral neck on the AP view.

Alternatively, these wires can be inserted through the plate with the locking guides attached.

Plate application

With the three wires in place, the plate with locking guides can be slid over the wires onto the side of the femur.

Screw insertion

The length of the screws is measured using a direct measuring device over the wire on the locking guide.

At this stage, a decision should be made whether a compression screw is required. If it is, it should be inserted through the third hole in the proximal part of the plate before any other screws are inserted. The compression screw can be a cortical lag screw. The hole for this screw should be drilled parallel to the 2.8 mm K-wires.

Note: A compression screw, if used, must be inserted before any locking screws.

After insertion of the compression screw, the locking screws should be inserted.

The guide wires are removed and replaced by locking screws one at a time. This ensures that fracture reduction is maintained by at least two wires, or screws, at all times.

If a compression screw has not been used, the third hole in the proximal part of the plate can be used to insert a further locking screw.

The plate should be secured to the shaft using two, or three locking screws. This prevents any disruption of the fracture fixation.

Definitive radiological images should be taken.

Closure

Consideration should be given to performing a capsulotomy for intracapsular fractures to prevent hemarthrosis under pressure.

The capsule should then be closed loosely to prevent a hemarthrosis forming under pressure, which can compromise the blood supply.

Routine closure is performed for the remainder of the wound, according to the surgeon’s preference.

4. Aftercare

Immobilization and weight bearing

If the fixation is stable, no external immobilization is required.

After a day or two of bed rest, the patient can be mobilized toe-touch weight bearing using crutches or a frame.

Increased weight bearing can be introduced as the fracture is judged to be healing depending on patient age. In an adolescent, partial weight bearing at 6 weeks and full weight bearing at 12 weeks would be typical.

Plate removal

If screws cross the physis, and a lot of growth potential remains, they should be removed when the fracture is healed and consolidated.

In small children, plate removal is advised to prevent bony overgrowth. Typically, this would be done at one year.

Plate removal is not compulsory in adolescents.