Executive Editor: Amy Kapatkin General Editor: Noel Moens

Authors: Bruno Peirone

Femoral shaft 32-C3

back to skeleton

Glossary

1 Indications top

enlarge

32-B3, 32-C3 and usually 32-B2 fractures are not reducible fractures; therefore, biological fracture repairs are recommended.

A plate with a bridging function can under certain conditions be used alone for internal fixation of these fractures. If a bone plate is used alone, it must be able to withstand all the weight bearing forces during healing, as the bone will not initially be sharing the load.
A bridging plate used alone is acceptable in light weight, young dogs in which bone healing is fast.
For larger patients, stronger fixation such as plate and rod or an interlocking should be used. Double plating could be considered as an alternative.

2 Principles top

enlarge

Fracture types:

A) Comminuted, unreduced fracture
B) Comminuted fracture aligned biologically 
C) Fracture aligned and biologically stabilized

3 Approach top

enlarge

OBDNT approach

An open-but-do-not-touch (OBDNT) approach provides direct visualization of the fracture site but the fracture fragments are minimally manipulated.


enlarge

MIPO approach

A Minimally Invasive Plate Osteosynthesis (MIPO) technique can be used. A surgical approach to the proximal and the distal femur is performed.

4 Surgical technique top

enlarge

Reduction

Indirect reduction is achieved by distracting and aligning the major bone segments using bone clamps, or other distraction techniques.


enlarge

Validation of alignment and rotation

Once the bone length has been restored, it is necessary to check for correct alignment and rotation. Rotational alignment can be judged by palpation or by direct visualization of the greater trochanter and femoral trochlea or by verifying the alignment of the
adductor magnus muscle. The distal part of the femur is held in a true lateral position. The position of the greater trochanter is then inspected.
If the femur is correctly aligned in the axial plane the greater trochanter should be slightly caudal to the long axis of the bone.


enlarge

The alignment can also be checked by confirming the orientation of the femoral neck relative to the plane of the femur.  This is done by inserting a small pin along side the femoral neck with the femur in a true lateral position. Orientation of the pin should be about 15°-25° in the cranial direction relative to the sagittal plane of the femur.


enlarge

Correct alignment and rotation can be checked with intraoperative imaging. The latero-medial projection is used and the whole bone, including the proximal and distal joints, must be visible.
If the femur is properly aligned in the transverse plane, about 1/3 of the femoral head should be visible cranially to the cranial femoral cortex. (Image A). Image B shows excessive internal rotation and C shows excessive external rotation of the distal femur.

Comparison with the contralateral unaffected limb can be useful.


enlarge

In a properly aligned leg, manual manipulation of the femur, will allow 90°  of external rotation and 45° of internal rotation of the hip. This method can only be used if the plate has been temporarily secured to the bone.


enlarge

Plate selection

A large plate must be selected because it must be able to withstand all the weight bearing forces because the bone will not be sharing the load. A lengthening plate (a plate without holes in the central part) can be used.


enlarge

Plate placement

The plate is contoured and placed on the lateral side of the femur, in bridging function.


enlarge

Plate application

If possible, the plate is applied by inserting at least three bicortical screws in each major segment. The plate should bridge at least 75% of the length of the femur.


enlarge

Locking plate

A locking plate can be used instead of a traditional bone plate.

5 Case example top

enlarge

Preoperative radiographs


enlarge

Postoperative radiographs (bridge plate)


enlarge

Postoperative radiographs at 1 month (failure bridge plate)


enlarge

Postoperative radiographs (Plate and rod)


enlarge

Postoperative radiographs at 2 months (revision rod and plate)

6 Alternative surgical technique: Double plating top

enlarge

Indications

Double plating can be an alternative to plate rod or interlocking nail fixations in some cases; for example, in very active, large patients or when a single plate is not considered to be strong enough.


enlarge

Application of plates

The first plate is applied on the lateral surface of the femur. A second plate is secured to the cranial surface.
During the cranial plate placement, care must be taken to avoid interference with the quadriceps mechanism. The distal plate end should be placed safely above the most proximal part of the trochlear ridge. Placement of the cranial plate and screw orientation should be carefully planned to minimize interference with the lateral plate and screws.


enlarge

1.0