Executive Editor: Chris Colton

Authors: Florian Gebhard, Phil Kregor, Chris Oliver

Distal femur 33-B3.1 ORIF

back to skeleton

Glossary

1 Principles top

Final osteosynthesis headless compression screw enlarge

General considerations

B3.1-type fractures may be associated with patellar dislocation/subluxation. These fractures may be such small osteochondral pieces of bone as to be technically unfixable. Small fracture fragments may need to be excised, or washed out of the knee joint.

Retropatellar surface
Check the retropatellar surface for lesions.

Pearl: decision making
In these injuries, there may be small osteochondral defects that are not reconstructable. Every reasonable attempt should be made to reconstruct the articular surface. On the other hand, a loose small fragment in the knee joint is clinically worse than a small articular defect. Therefore, surgical judgment must be exercised in determining whether a fragment is reconstructable, or not.

Specialist knee reconstruction surgeons have available advanced techniques to reconstruct the articular surface. Osteochondral autograft and allograft transfer within the knee are possible. Cultured chondrocyte grafting shows promise for the future.

2 Reduction and temporary fixation top

enlarge

Reduction

Achieve reduction by manipulating the fracture with an elevator. Small ball-spiked pushers, or a dental pick, may also be used.


Achieve reduction using arthroscopic control enlarge

Alternative: reduction under arthroscopic view

Achieve reduction by manipulating the fracture with the hook (illustrated) using arthroscopic control. Small ball-spiked pushers may also be used, inserted through separate portals.


Inserting a K-wire enlarge

Temporary fixation

Insert a K-wire to secure the reduction. Make sure that it does not interfere with the intended implant position.

3 Fixation top

Various types of fixation enlarge

General consideration

Fixation can be achieved using a small fragment screw system (A or B), or absorbable pins (C). In general screws provide a better compression of the fracture, but need a certain thickness of the osteochondral flake due to the screw head dimension.

Absorbable pins (C) are preferable for thin and small fragments.

Two screws, or pins, should be used in order to prevent fragment rotation.

None of the implants is allowed to project above the articular surface.


Placing the guide wires to the fracture plane enlarge

Guide-wire placement

Place the guide wires as perpendicularly as possible to the fracture plane.

Check correct guide-wire placement under image intensifier control.


Countersinking enlarge

Countersinking

Countersink manually, deep enough to allow for fully buried screw heads.

Pitfall: countersinking too deep
Be aware of countersinking too deeply into the cancellous bone, which often happens using a power tool.


Sunken screws enlarge

Sunken screws

In screw fixation the lag screw principle is followed. In B3.1-type fractures use partially threaded lag screws so that no gliding hole need to be drilled. It is imperative to countersink for the screw head.


Completed osteosynthesis using sunken screws enlarge

This illustration shows the completed osteosynthesis using sunken screws.


Headless compression screws inserted enlarge

Headless compression screw

Insert 3.0 mm headless compression screws with predrilling, or over a guide wire. Neither gliding hole nor countersinking is required. It is imperative to insert the headless compression screw until the screw head has passed the cartilage surface completely.


Completed osteosynthesis using headless compression screws enlarge

This illustration shows the completed osteosynthesis using headless compression screws.


Absorbable pins inserted enlarge

Absorbable pins

Absorbable pins require predrilling. Use the insertion device to insert the pin with gentle hammer blows. Observe the pin insertion process carefully as the pins might deform during insertion.

In case the pin head does not completely sink into the cartilage use a scalpel to remove the projecting part of the head.


Completed osteosynthesis using absorbable pins enlarge

This illustration shows the completed osteosynthesis using absorbable pins.

v1.0 2008-12-03