Executive Editor: Chris Colton

Authors: Florian Gebhard, Phil Kregor, Chris Oliver

Distal femur 33-B3.2/3 ORIF

back to skeleton

Glossary

1 Principles top

Completed osteosynthesis enlarge

Fixation can be achieved indirectly with 3.5 mm or 4.5 mm screw systems. It is advantageous to use cannulated screw systems.

In small fragments direct fixation is required through a posterior approach. Click here for a description of screw fixation from posterior.

At least two screws should be used, in order to prevent fragment rotation and to achieve satisfactory compression.

No implant is allowed to violate on the articular surface.

Similar principles apply if both condyles are fractured (B3.3-type fracture).

2 Reduction top

The joystick technique enlarge

Achieve reduction using periosteal elevators and a large pointed reduction forceps. The joystick technique, in which a small Schanz screw is inserted from the extraarticular surface, is also useful (as illustrated).

3 Fixation top

Inserting guide wires enlarge

Guide-wire insertion

Insert the guide wires as perpendicularly as possible to the fracture plane.


The condylar surface must not be perforated enlarge

Check guide-wire placement

Check the correct guide-wire insertion under fluoroscopic image intensifier control. The condylar surface must not be perforated.


Countersinking manually enlarge

Countersinking

Countersink manually, deeply enough to allow for fully buried screw heads.

Pitfall: countersinking too deeply
Be aware of countersinking too deeply into the cancellous bone, which often happens using a power tool.

Do not perform countersinking when planning the use of headless compression screws.


Determining the appropriate screw lengths enlarge

Screw length determination

Determine the appropriate screw lengths, inserting the dedicated measuring device into the countersunk hole.


Inserting the screws enlarge

Screw insertion

Insert the screws using the cannulated screw driver and check under image intensifier control in the lateral and oblique views that the screw length was chosen correctly.


X-ray showing the completed osteosynthesis enlarge

X-ray

X-ray showing the completed osteosynthesis.

v1.0 2016-10-19