Executive Editor: Chris Colton, Steve Krikler

Authors: Pol Rommens, Peter Trafton, Martin Jaeger

Humeral shaft - Multifragmentary fracture, fragmentary segment

back to skeleton

Glossary

1 Principles top

Plating of proximal 1/3 fractures enlarge

Plating of proximal 1/3 fractures

In fractures with markedly displaced proximal, metaphyseal fragmentation it is crucial to still achieve anatomic or almost anatomic reduction. The older the patient, the more important this is, to prevent nonunion and regain optimal function.

To achieve this, the plate should provide sufficient fixation points in the proximal humerus/humeral head. It also should be long enough. Typically, 6 cortical fixation points distal to the fracture zone are sufficient.

Bridge plating alone does not address grossly displaced fragments. In the elderly patient this may lead to delayed or nonunion. Therefore, consider adding techniques such as cerclage or lag screw fixation of large fragments, or choose a different treatment.


Deltopectoral approach enlarge

Approach

Depending on the proximal fracture morphology and surgeon’s preference a deltopectoral, extended anterolateral or lateral approach may be used in a beach chair or supine position.

In this procedure, the deltopectoral approach is shown. The principles of fixation remain the same regardless of the approach used.


Choice of plate enlarge

Choice of plate

3.5 mm locking plate systems, such as PHILOS-Long are suitable and shown in this procedure.

2 Reduction and lag screw fixation top

Reduction

It is important to restore axial alignment and rotation.

In osteoporotic bone anatomic reduction should be achieved as callus formation is poor.

Anatomic reduction and internal fixation follows the principle of reducing the numbers of fragments from 4 to 3, 3 to 2 and 2 to 1 starting distally.

Alternatively, a bridging technique may be used.


Lag screw fixation enlarge

Lag screw fixation

If the fragments are big enough and the fracture planes are suitable, the fragments may be reduced and held with lag screws or cerclage.

Reduce the most distal fragment to the distal humerus, hold it with a pointed reduction forceps and fix it with a lag screw.


Lag screw fixation enlarge

Then proceed with the next fragment and fix it to the proximal humerus.


Lag screw fixation enlarge

If possible reduce the two main fragments using a pointed reduction forceps. Take care of proper rotation.

Option: A further lag screw may be used but typically it is not necessary. 

3 Plate fixation top

Plating of proximal 1/3 fractures enlarge

Plate position

The plate may be applied either in the lateral or the anterolateral position.

Lateral position
For the lateral position the plate has to be passed through a funnel under the deltoid muscle.


Plating of proximal 1/3 fractures enlarge

Anterolateral position
The aim of the anterolateral position is to protect the deltoid insertion as much as possible and not to interfere with the course of the long head of the biceps.

The proximal portion of the plate is placed on the lateral surface and the distal portion is placed on the anterior surface of the humerus. This requires the plate to be twisted to form a helix, with sufficient twist to allow the plate to pass anteriorly to the deltoid insertion.


Plating of proximal 1/3 fractures enlarge

Plate contouring

For anterolateral plate positioning twist the distal end of the plate internally to match the orientation of the anterolateral surface of the distal humerus.

A malleable template is helpful for matching contours of proximal and distal segments. Alternatively, a rod can be used to estimate the amount of twisting.


Plating of proximal 1/3 fractures enlarge

Plate application

Abduct the arm to release the tension of the deltoid. If the stability of the initial fixation is adequate, slight internal rotation and retrotorsion of the arm makes proximal access easier.

If the anterior proximal deltoid insertion still interferes with the plate position it may be released.


Plating of proximal 1/3 fractures enlarge

Take care not to interfere with the long head of biceps.

If the plate has been twisted the distal end of the plate should lie in an interneurovascular split of the brachialis muscle.


Confirm the plate position by palpation and under image intensification. enlarge

Confirm the plate position by palpation and under image intensification. Check that the plate is:

  • Not positioned too high (risk of impingement)
  • Nicely aligned to the axis of the humerus
  • Providing sufficient fixation points distal to the fracture zone

Plate fixation to humeral head enlarge

Plate fixation to humeral head

The number of screws depends on the bone quality. In osteoporotic bone a higher number of screws may be required.


Start with a bicortical nonlocking screw through the elongated hole of the PHILOS in the meta/diaphyseal area. enlarge

If possible, start with a bicortical nonlocking screw through the elongated hole of the PHILOS in the meta/diaphyseal area. This allows fine tuning of the plate position.


Use the guide block with drill sleeves to drill holes for the proximal screws. enlarge

Drilling hole
Use the guide block with drill sleeves to drill holes for the proximal screws. Do not drill through the subchondral bone and into the shoulder joint.

Follow the manufacturer’s technique guide.

Two drilling techniques help to avoid drilling into the joint.

  • “Woodpecker”-drilling technique (as illustrated): Advance the drill bit only for a short distance, then pull the drill back before advancing again. Keep repeating this procedure until subchondral bone contact can be felt. Take great care to avoid penetration of the humeral head.
  • Particularly in osteoporotic bone, one can drill only through the near cortex. Push the depth gauge through the remaining bone until subchondral resistance is felt.

Determine screw length. enlarge

Determine screw length
The intact subchondral bone should be felt with an appropriate depth gauge or blunt pin to ensure that the screw stays within the humeral head. The integrity of the subchondral bone can be confirmed by palpation or the sound of the instrument tapping against it. Typically, choose a screw slightly shorter than the measured length.


Insert screw. enlarge

Insert screw
Insert a locking head screw through the screw sleeve into the humeral head. The sleeve aims the screw correctly. Particularly in osteoporotic bone, a screw may not follow the hole that has been drilled. 


Place a sufficient number of screws into the humeral head. enlarge

Place a sufficient number of screws into the humeral head.


Plate fixation to humeral shaft enlarge

Plate fixation to humeral shaft

Distally locking screw fixation is preferred. Again, the number of cortical fixation point depends on the fracture morphology and bone quality.

Insert at least three screws into the humeral shaft, distal to the fracture zone, with appropriate, percutaneously placed screw insertion sleeves.


Completed osteosynthesis enlarge

This illustration shows the completed osteosynthesis.

4 Final check of osteosynthesis top

Final check of osteosynthesis enlarge

Using image intensification, carefully check for correct reduction and fixation (including proper implant position and length) in various arm positions. Ensure that screw tips are not intraarticular.


X-ray of osteosynthesis with PHILOS and lag screws enlarge

X-ray of osteosynthesis with PHILOS and lag screws

v2.0 2018-12-28