Executive Editor: Chris Colton, Steve Krikler

Authors: Pol Rommens, Peter Trafton, Martin Jaeger

Humeral shaft - Multifragmentary fracture, intact segment

back to skeleton

Glossary

1 Principles top

Straight humeral nail enlarge

General considerations

Acute and pathological fractures can be nailed. The best indications for humeral shaft nailing are fractures located between the surgical neck and approximally 5 cm above the olecranon fossa.

There are many different nails on the market; the main difference between them relates to their nail design (straight vs bent nails) and their locking configurations. Some nails allow compression at the fracture site.

To demonstrate the principles, the illustrations below show the use of a straight nail. Some details may not be applicable with other nailing systems.

For discussion of bent vs straight nails see the corresponding additional material.


In fractures with markedly displaced proximal, metaphyseal fragmentation it is crucial to still achieve anatomic or almost anatomic reduction. enlarge

Reduction and approach

In fractures with markedly displaced proximal, metaphyseal fragmentation it is crucial to still achieve anatomic or almost anatomic reduction. The older the patient, the more important this is, to prevent nonunion and regain optimal function.


Deltopectoral approach enlarge

Depending on the proximal fracture morphology and surgeon's preference it may then become necessary to use a deltopectoral, extended anterolateral or lateral approach. Remember that reduction of the proximal fragments is necessary prior inserting the nail.

In this procedure, the deltopectoral approach is shown. The principles of fixation remain the same regardless of the approach used.


Beach chair position enlarge

Patient positioning

For antegrade nailing, the beach chair is the preferred position.

If no beach chair is available, a supine position may be used.


Interlocking

Multifragmentary fractures cannot be fixed with compression. Therefore, in general, static interlocking with at least two screws on each side of the fracture is recommended. Depending on the fracture morphology/extension and the bone quality, additional locking screws may be used.

Interlocking is performed from proximal to distal.

2 Determining nail diameter and length top

Nail diameter enlarge

Nail diameter

Place a radiographic ruler anteriorly on the upper arm over the medullary canal.

Visualize the medullary canal under image intensification in AP view and compare with the diameters of the marks on the ruler.

The correct diameter of the humeral nail corresponds to the diameter of the mark that fits completely into the medullary canal. Look carefully at the distal humeral canal. It may be narrower than the rest. If the isthmus is very narrow, consider reaming to allow the insertion of a larger diameter nail.

In our example, a nail of 7.5 mm diameter is chosen.

The diameter of the medullary cavity can also be measured on digital radiographs of the humerus. Take care to estimate the diameter on the narrowest site.


Nail length enlarge

Nail length

Provisionally reduce the humeral fracture.

Place the ruler parallel to the arm, with the top-of-nail mark at the level of the anticipated insertion site in the humeral head.

Take an AP image intensifier view of the distal humerus together with the ruler.

The humeral nail should be as long as possible in the distal canal, consistent with its diameter. Determine this length with the aid of the ruler.

Keep in mind that the top of the nail should lie just below the surface of the bone.

The length of the nail can also be estimated by measurements of digital preop radiographs. Care should be taken that the whole humerus is pictured.

3 Reduction and fixation of proximal fragments top

Open reduction enlarge

Open reduction

Reduce the fracture under direct vision, with the aid of blunt hooks or reduction forceps if necessary.


Cerclage enlarge

Cerclage

Bicortical lag screws may interfere with the passage of the nail and so should not be used. Therefore, cerclage is recommended for fixation.

Place a cerclage wire using a wire passer around the humerus with continuous direct contact to the bone to avoid neurovascular injuries.

If in any doubt do not hesitate to confirm the passage either by palpation or under direct vision.

4 Entry point top

Location of entry point enlarge

Location of entry point

Different nailing systems have nails of different shapes, and so may require different entry points. Refer to the manufacturers' operative technique guides for further details.

The insertion point for the straight nail is in line with the medullary cavity. Typically, it ends up at the highest point of the humeral head. In contrast to bent nails it is located intraarticularly completely, well away from the insertion of the supraspinatus tendon.


Insertion of guide wire enlarge

Insertion of guide wire

Insert a guide wire through the correct entry point and confirm proper placement by image intensification.


Opening the canal enlarge

Opening the canal

A cannulated drill bit is recommended for opening the proximal humerus.

Insert this drill bit over the guide wire and advance into the proximal medullary canal.


Opening the canal enlarge

Alternatively, use an awl to open the humerus.

5 Reaming top

Indication

If the canal is very narrow rather than insert a narrow nail, reaming will allow for the insertion of a larger diameter (so stronger) nail. This may be particularly useful when nailing a nonunion.


Reaming enlarge

Procedure

Pass an olive tipped guide wire across the fracture site. With experience it is usually possible to feel the tip of the wire is in the cancellous bone, but this should be confirmed with image intensifier.


enlarge

Ream up sequentially to an appropriate size for the desired nail. In segmental or multifragmentary fractures take care to avoid soft-tissue damage, particularly to the nerves and vessels, as the reamer is passed through the fracture zone.

Exchange the olive tipped wire to a straight guide wire, and then insert the nail over the guide wire. Alternatively, some nails may be introduced without a guide wire.

6 Nail insertion top

Insertion handle enlarge

Insertion handle

Mount the humeral nail on the insertion handle. The nail must be rotated correctly relative to the insertion handle.


Nail insertion enlarge

Nail insertion

Insert the nail with slightly rotating movements down to the fracture line.

Do not use a hammer, as this may increase the risk of fissure or fracture at the insertion site.


Advance the nail under image intensification just beyond the fractures. enlarge

Advance the nail under image intensification just beyond the fractures.

If inserting an unreamed nail a guide wire may be helpful but may not be necessary.


Passing the nail into the distal fragment enlarge

Passing the nail into the distal fragment

Pass the fracture zone under image intensification and make sure that the nail enters the distal fragment properly.

Use the nail tip as a reduction aid. Alternatively, it may be easier to pass a guide wire with a bent tip across the fracture zone, and then pass the nail over the guide wire.

After passing the fracture zone, adjust humeral shaft alignment, rotation, and length.

Confirm with image intensification in two different planes.


Definitive nail position enlarge

Definitive nail position

Gently advance the nail until it has reached the desired position just above the olecranon fossa. Avoid fracture distraction.

If a guide wire was used, remove it once the nail is in place.


Make sure the proximal end of the nail is placed beneath the bony surface of the humeral head. enlarge

Make sure the proximal end of the nail is placed beneath the bony surface of the humeral head.

No protrusion of the nail may be tolerated. Confirm with appropriately oriented C-arm images that the nail is below the bone.


A K-wire placed through an aiming device may help to adjust the appropriate insertion depth. enlarge

A K-wire placed through an aiming device may help to adjust the appropriate insertion depth.


Rotational reduction enlarge

Rotational reduction

Confirm the fracture reduction under image intensification. Ideally the fracture line should be closed anatomically.

In cases of comminution indirect signs can be helpful such as the width of the cortices proximal and distal to the fracture site. Both should have the same width. Any difference suggests rotational malalignment.

Alternatively, if the proximal fragment is in the correct anatomic position, confirmed with image intensification, the forearm should lie in 25° external rotation.


Rotational implant position enlarge

Rotational implant position

To lock the nail in the correct rotation, swivel the aiming arm approximately 25° anteriorly. Due to the physiological retrotorsion of the humeral head, the axis of the humeral head is directed approximately 25° posteriorly to the condylar plane of the distal humerus.


Rotational implant position enlarge

To obtain a true AP view of the proximal humerus, rotate the forearm approximately 25° externally relative to the sagittal plane.


Pitfall: interference of the long head of the biceps enlarge

Pitfalls

To achieve correct locking of the nail and avoid complications it is crucial to obtain the correct rotation of the nail in the humerus.

Typical problems arising after incorrect rotation of the nail are:

  • Proximally: interference of the long head of the biceps

Pitfall: interference of the screw insertion with neurovascular structures enlarge
  • Distally: interference of the screw insertion with neurovascular structures

Longitudinal reduction enlarge

Longitudinal reduction

Confirm reduction of the fracture under image intensification. Take care that there is no fracture gap as this has a high risk of nonunion.

If there is an excessive gap, it may be possible to close it with manual compression along the axis of the humerus, telescoping the distal fragment over the nail.


Longitudinal reduction enlarge

Alternatively, the gap can be closed with retrograde pulling of the nail after initial distal locking.

This requires the nail to be inserted deeper into the proximal fragment, so that as the fracture gap is closed the top of the nail comes to lie in the correct position. Otherwise an intraarticular, proximal protrusion of the nail will cause injury to the rotator cuff and impingement.

7 Proximal interlocking top

Trocar insertion enlarge

Trocar insertion

Place the trocar for the planned screw fixation through the aiming device, if necessary through a separate skin incision. Perform blunt dissection down to the bone so that the trocar may be placed without interposition of soft tissue.


Predrilling enlarge

Predrilling

Check once more the correct retrotorsion.

Drill with the calibrated drill bit carefully just to the level of the subchondral bone. The final position of the drill bit can be checked with the image intensifier. The screw length can be determined from the measurements of the drill bit.


Screw hole widening in the lateral cortex enlarge

Screw hole widening in the lateral cortex

If the proximal locking bolts have enlarged heads the lateral hole must be overdrilled with the appropriate drill to avoid fracturing the lateral cortex.


Screw insertion enlarge

Screw insertion

Attach the screw to the screw driver and insert the screw.

In some systems, as the one shown in this illustration, a screw-in-screw fixation is possible (see below) and may help in fractures involving the humeral head. If so and if also intended, the plane of the handle of the screw driver should be horizontal and the arrow should point posteriorly once the final position of the screw is reached.


Insertion of further screws enlarge

Insertion of further screws

Insert further screws using the aiming device according to the fracture pattern.

Remove the aiming device.

8 Distal interlocking top

Position of the arm enlarge

Position of the arm

Distal interlocking is typically performed using a freehand technique. Therefore, the arm has to be placed in a stable position for distal locking under image intensification.


Defining the correct position of the arm/C-arm enlarge

Defining the correct position of the arm/C-arm

It is crucial for a proper freehand interlocking to define the correct position of the arm/C-arm. This is defined under image intensification showing a perfectly round circle of the nail's locking holes within the center of the image.


Pitfall: unrecognized malrotation enlarge

Pitfall: unrecognized malrotation

While positioning the arm and determining the proper position of the C-arm pay special attention not to produce an unrecognized malrotation by just rotating the forearm.


Skin incision enlarge

Skin incision

A rare but serious complication is injury to either the brachial artery and/or the median nerve while inserting the distal locking screw.

Therefore, take special care, using blunt dissection. If in any doubt widen the exposure and use retractors to protect the neurovascular structures.


Skin incision enlarge

Position the tip of the scalpel over the proximal hole, as shown under image intensification, and incise the skin.


Soft-tissue dissection enlarge

Soft-tissue dissection

Use small scissors or artery forceps to prepare a track for insertion of the drill bit.


Drilling enlarge

Drilling

Perform the positioning of the drill tip, its fine adjustment, and the drilling procedure under image intensification.

Many surgeons prefer a radiolucent drill attachment if it is available.


Center the radiopaque circles of the radiolucent drill on the locking hole. enlarge

Center the radiopaque circles of the radiolucent drill on the locking hole.

Measure the length of the required screw with a depth gauge.


Insertion of interlocking screws enlarge

Insertion of interlocking screws

Insert the first interlocking screw and check for the correct position and length under image intensification in two planes.

Insert further interlocking screws as described above.


Pearl: absorbable suture around screw enlarge

Pearl: absorbable suture around screw

To avoid losing the screw in the soft tissues during insertion and tightening, fasten an absorbable suture around the screw’s head beforehand.


When interlocking is completed, check the definitive position of the implant under image intensification in two planes. enlarge

Final remark

When interlocking is completed, check the definitive position of the implant under image intensification in two planes.


Take particular notice of the position of the proximal end of the nail, the location and length of all interlocking screws. enlarge

Take particular notice of the position of the proximal end of the nail, the location and length of all interlocking screws.

9 Nail capping top

Nail capping enlarge

Remove the insertion handle/aiming guide and its connecting screw.

Insert an end cap to prevent ingrowth of bony tissue.


Nail capping enlarge

End caps of different lengths may be available. The top of the end cap should lie just below the articular cartilage, to facilitate later implant removal, but must not protrude above the humeral surface.

Check the proper dimension of the end cap under image intensification.

v2.0 2018-12-28