Executive Editor: Fergal Monsell General Editor: Chris Colton

Authors: Andrew Howard, James Hunter, Theddy Slongo

Pediatric proximal femur 31-M/3.1 III

back to Pediatric overview

Glossary

1 Preliminary remarks top

enlarge

Introduction

Children's hip fractures have historically had a bad prognosis. Accurate stable fixation has been one of the factors identified to improve the outcome.

The main problems have been avascular necrosis (AVN) due to disruption of the blood supply, nonunion, and malunion due to unstable fixation.

Extracapsular fractures have a reduced incidence of AVN, but are prone to malunion without stable fixation.

The objective is timely fixation without disruption of the blood supply.

Angularly stable fixation is the solution to malunion. This is the rationale for using a pediatric hip locking plate.


enlarge

Implant selection

The choice of implant size depends on the age and weight of the child.

Generally, a 130° angled plate is appropriate for fracture fixation as it corresponds most closely to the child's femoral neck-shaft angle. It offers the possibility of three parallel screws into the neck.

Plate length depends on the extension of the fracture down the femoral shaft. The pediatric hip locking plate has a choice of three, five, seven, and nine holes.

Longer locking screws, not included in the standard pediatric hip locking plate set, may be required.


enlarge

Implant placement

During preoperative planning, a decision should be made whether fixation needs to cross the physis. This decision will depend on the level of the fracture.

For extracapsular fractures, crossing the physis is less likely to be required.

2 Reduction top

enlarge

Reduction is achieved by a combination of gentle traction, leg position, and direct reduction.


enlarge

In larger children and adolescents, a fracture table may aid reduction, particularly if a surgeon has limited assistance.


enlarge

If reduction is incomplete after traction and correction of rotation, direct reduction with a clamp may be necessary.


enlarge

Temporary K-wire stabilization with two divergent wires is useful at this stage. Care must be taken that the temporary K-wire(s) does not prevent insertion of the definitive wires.

3 Fixation top

enlarge

Insertion of guide wires

After reduction, the guide wires for the pediatric hip locking plate are inserted.

This can be done using the dedicated angled guide, or by placing the wires through the plate with the screw-in guides attached to it.

An anteversion wire along the front of the femoral neck can be useful.


enlarge

A 2.0 mm guide wire should be inserted high in the neck starting at least 5 mm below the trochanteric apophysis. This wire can be inserted using the angled guide or freehand.

The position should be checked radiologically. On the lateral view, the wire should be in the middle of the femoral neck.

After insertion of the initial guide wire for the pediatric hip locking plate, it may be necessary to remove one or more temporary K-wires to facilitate use of the angled guide.


enlarge

After the initial 2.0 mm wire, two 2.8 mm wires will be inserted using the angled guide.

The 2.8 mm wires will sit just above the mid-point of the femoral neck on the AP view.


enlarge

Alternatively, these wires can be inserted through the plate with the locking guides attached.


enlarge

Plate application

With the three wires in place, the plate with locking guides can be slid over the wires onto the side of the femur.


enlarge

Screw insertion

The length of the screws is measured using a direct measuring device over the wire on the locking guide.


enlarge

At this stage, a decision should be made whether a compression screw is required. If it is, it should be inserted through the third hole in the proximal part of the plate before any other screws are inserted. The compression screw can be a cortical lag screw. The hole for this screw should be drilled parallel to the 2.8 mm K-wires.

Note: A compression screw, if used, must be inserted before any locking screws.


enlarge

After insertion of the compression screw, the locking screws should be inserted.

The guide-wires are removed and replaced by locking screws one at a time. This ensures that fracture reduction is maintained by at least two wires, or screws, at all times.

The shaft fragment may be controlled using a loosely applied bone-plate clamp or a partially tightened monocortical screw.  Care should be taken not to disrupt the fracture reduction by overtightening the plate to the shaft.


enlarge

If a compression screw has not been used, the third hole in the proximal part of the plate can be used to insert a further locking screw.


enlarge

The plate should be secured to the shaft using an appropriate number of locking screws. This prevents any disruption of the fracture fixation. The bone-plate clamp is removed.

Definitive radiological images should be taken.


Closure

Routine closure of the wound is performed, according to the surgeon’s preference

4 Aftercare top

enlarge

Immobilization and weight bearing

If the fixation is stable, no external immobilization is required.

After a day or two of bed rest, the patient can be mobilized toe-touch weight bearing using crutches or a frame.

Increased weight bearing can be introduced as the fracture is judged to be healing, depending on patient age. In an adolescent, partial weight bearing at 6 weeks and full weight bearing at 12 weeks would be typical.


Plate removal

If screws cross the physis, and a lot of growth potential remains, it should be removed when the fracture is healed and consolidated.

In small children, plate removal is advised to prevent bony overgrowth. Typically, this would be done at one year.

Plate removal is not compulsory in adolescents.

v1.0 2017-12-04