Distal forearm 23-A2.2 ORIF

back to skeleton

Glossary

1 Preliminary remarks top

enlarge

Fracture assessment and decision making

In A2.2 type fractures, there is a fracture of the distal metaphysis with dorsal displacement. These are often referred to as Pouteau or Colles fractures.

Fractures angulated dorsally >25° associated with osteoporosis or residual void after reduction may prove unstable. Therefore, primary palmar plating may be chosen.

Before palmar plating became a commonly used treatment, most of these fractures were treated with closed reduction, which was then maintained either with K-wires or a plaster cast. Many surgeons now treat the majority of these fractures with a palmar plate and often use the plate as an aid to reduction.


enlarge

Palmar plate

Advances in plate design have provided angular stable fixation. This allows enhanced stability and ease of application, even in the presence of osteoporotic bone. Plates with variable angle locking screw options may be useful.


enlarge

Provisional reduction in displaced fractures

Reduction is achieved by applying longitudinal traction either manually or using Chinese finger traps.

The reduction is maintained by a temporary splint.

If definitive surgery is planned, but cannot be performed within a reasonable time scale, a temporary external fixator may be helpful.

2 Associated injuries top

enlarge

Median nerve decompression

If there is dense sensory loss, or other signs of median nerve compression, the median nerve should be decompressed.

3 Reduction and fixation top

enlarge

Insert K-wire

Insert a small smooth K-wire or needle to mark the level and plane of the radiocarpal joint.

If there are fragments blocking reduction they may be removed with a dental pick.


enlarge

Reduction with plate

Apply the plate to the distal fragment. The distal end of the plate should end at the anatomic watershed zone of the distal radius.

Insert a K-wire through a screw hole, as close to the subchondral bone as possible and parallel to the articular surface.

The angle of the plate to the shaft should equal the angle of the displacement.

Confirm using image intensification.


enlarge

The initial screw is inserted in the most ulnar screw hole. The reason for this is that if the initial screw is placed on the radial side it will block accurate imaging of the ulnar screw placement. Choose a locking head screw, 2-4 mm shorter than measured.

Provided the screw is parallel to the K-wire, it should not enter the radiocarpal joint.


enlarge

Confirm screw position with a lateral view under image intensification, with the beam aimed at an angle of 20° to the true lateral, clearly showing the joint surface.


enlarge

Insert at least two other distal locking head screws.


enlarge

Pitfall: Screw tip protrusion

Due to the prominence of Lister’s tubercle, as seen on the lateral image projection, a screw placed on either side of the tubercle may appear not to protrude through the far cortex.

Protrusion of such a screw may result in extensor tendon irritation and rupture.


enlarge

Bring the plate onto the shaft and hold it with a clamp. Obtain check radiographs and adjust the position of the distal fragment if necessary by moving the plate.


enlarge

Once satisfactory reduction is confirmed, insert a self-tapping, non-locking screw through the oblong plate hole.


enlarge

Insert at least two further proximal screws.


enlarge

Alternative reduction

Some surgeons feel that using the plate for reduction in osteoporotic patients may cause the screws to loosen in the bone. In such cases, manual reduction and preliminary fixation with K-wires may be preferable.


enlarge

Apply the plate to the bone. The distal end of the plate should end at the anatomic watershed zone of the distal radius.

Insert a screw through an oblong hole in the proximal radial fragment. Before fully tightening it, check the plate position using intraoperative imaging, adjusting the position of the plate as necessary.