Executive Editor: Peter Trafton

Authors: Kodi Kojima, Steve Velkes

Proximal forearm 21-A2.3 Arthroplasty

back to skeleton

Glossary

1 General considerations top

enlarge

Indications

Stable restoration of radial length is usually important for elbow or forearm stability. If this cannot be achieved with ORIF, a proximal radius prosthesis will be necessary.

Prosthetic replacement for unreconstructable radial neck fractures is indicated:

  • When the elbow joint is unstable
  • With an unstable coronoid fracture
  • With medial collateral ligament insufficiency or ulnohumeral instability
  • With associated interosseous membrane injury (Essex-Lopresti injury)

Correct prosthesis size

Select the right prosthesis size to avoid over- or underfilling of the radiocapitellar joint which can cause restricted range of motion (too large and/or long prosthesis) or elbow instability (too small and/or short prosthesis). For comminuted radial neck fractures, determining the correct length of prosthesis may be difficult. Comparison with an x-ray of the opposite (intact) elbow may be helpful for preoperative planning.

Pearl:
Reconstruct the radial head and neck with the excised fragments and choose the size of the prosthesis accordingly.


Cemented or uncemented prosthesis

The surgeon must choose between cemented and uncemented prosthesis, the latter being easier, but perhaps less stable.

2 Resection of the proximal radius top

enlarge

Resect radial head and neck fragments

Carefully resect all radial head and neck fragments. Preserve the annular ligament for repair, if possible.


enlarge

Trim the radial neck

Trim the radial neck to fit the prosthesis with of a small rongeur.

3 Replacement top

enlarge

Opening the medullary canal

Carefully open the medullary canal with an awl to fit the prosthesis stem.


enlarge

Trial insertion of the prosthesis

Insert the chosen prosthesis. Assess its length and stability. Cementing the prosthesis may be necessary for optimal stability, and can be determined now.


enlarge

Avoid lengthening or shortening

To avoid lengthening and overstuffing of the radiocapitellar joint, or shortening and instability, the prosthesis should fit as follows:

The articular surface of the radial head prosthesis should be at the level of, or slightly proximal to the lateral edge of the coronoid articular surface.

The radius with prosthesis should match radiographs of the opposite (intact) forearm, to ensure correct length.


Assessment

Check tracking of the prosthesis in flexion, extension, pronation and supination.

Check elbow stability.

If the elbow is stiff or unstable, change the size of the prosthesis accordingly.

Cement the prosthesis in place, if desired, and if its size and position are satisfactory.

If the annular ligament is ruptured, repair it with non-absorbable sutures.

v1.0 2016-10-24