Executive Editor: Chris Colton

Authors: Dominik Heim, Shai Luria, Rami Mosheiff, Yoram Weil

Forearm shaft 22-A2.1 Lag screw and plate fixation

back to skeleton

Glossary

1 Principles top

Options for fixation

For oblique forearm shaft fractures, interfragmentary compression can be achieved either by using an independent primary lag screw, protected by a plate (A), or by using a compression plate as the primary loading device with the addition of a subsequent lag screw through the plate (B).

Note: The first principle (A) will be described using two different techniques.


Principles enlarge

Absolute stability

Absolute fracture stability, which is defined as the complete abolition of interfragmentary movement, is achieved by interfragmentary compression. It results in direct bone healing (in simple fracture configurations).

Interfragmentary compression
Interfragmentary compression can be achieved by a lag screw, inserted either through a plate hole (1), or separate from the plate (2).


Principles enlarge

If available, a partially threaded cortical lag screw (shaft screw) can be used to accomplish interfragmentary compression.


Principles enlarge

Dynamic compression principle

The holes of the plate are shaped like an angled cylinder. The spherical undersurface of the screw head slides down the inclined cylinder as the screw is tightened.


Principles enlarge

The horizontal movement of the head, as it impacts against the angled side of the hole, results in movement of the bone fragment relative to the plate, and leads to compression of the fracture.

Axial compression
Interfragmentary compression can be achieved by loading the oblique fracture site axially, using specific techniques.
Axial compression, using self-compressing plates (DCP, LC-DCP, LCP, etc.), is achieved by eccentric screw placement.


Principles enlarge

Plate position

Depending on the procedure chosen, the plate will be applied to either the anterior or posterior surface of the radius.

In the following example, we illustrate the plate applied to the posterior surface.


For proximal radial shaft fractures, the anterior approach (Henry) is most often used to minimize the risk of damage to the posterior interosseous nerve, which crosses the proximal radius within the supinator muscle.

In mid and distal radial shaft fractures, either the anterior approach (Henry) or posterolateral approach (Thompson) can be used, depending on surgeon’s preference.

2 Reduction top

Reduction enlarge

Open and anatomical reduction

Reduce the fracture anatomically, using a reduction forceps on each main fragment. The use of blunt, as opposed to pointed, reduction forceps can be helpful, particularly if greater forces are required.


Reduction enlarge

Pearl: leverage technique for oblique fractures
A small bone lever can be used to reduce the fragments.


Reduction enlarge

Pearl: lengthening by twisting of reduction forceps
Reduction of overlapping oblique fractures can sometimes be achieved by twisting a reduction forceps, thereby lengthening the fracture.


Reduction enlarge

Maintain fracture reduction

Maintain the fracture reduction with pointed reduction forceps. Place the forceps such that it will not interfere with the planned plate position.

3 Plate length and number of screws top

Plate length and number of screws enlarge

In the forearm, in addition to any lag screw, three bicortical plate screws are required in each main fracture fragment due to the high torsional stresses. For biomechanical reasons, not every plate hole needs to be occupied by a screw. Therefore, plates with at least 8 or 9 holes are usually used.

4 Fixation – lag screw as primary fixation device separate from plate top

Fixation – lag screw as primary fixation device separate from plate enlarge

If the fracture morphology dictates that the interfragmentary lag screw be positioned separate from the plate, the plate then acts as a neutralization (protection) plate and axial compression should not be applied.


Fixation – lag screw as primary fixation device separate from plate enlarge

Drill gliding hole

Using a 3.5 mm drill sleeve and a 3.5 mm drill bit, drill a gliding hole in the near cortex.
Ensure that the direction of the drill is as perpendicular to the fracture plane as possible.


Fixation – lag screw as primary fixation device separate from plate enlarge

Thread hole: guide in gliding hole

Insert the 3.5 mm / 2.5 mm drill guide into the gliding hole. Check that the anatomical reduction is maintained and use a 2.5 mm drill bit to drill a pilot hole just through the far cortex.


Fixation – lag screw as primary fixation device separate from plate enlarge

Measure for screw length

Countersink the cortex at the gliding hole.
Use a depth gauge to measure for screw length.
Measure the longer side of the oblique drill hole, as shown, to ensure sufficient screw length.
The lag screw should protrude 1-2 mm through the opposite cortex to ensure maximal thread purchase. However, too long a screw may be tender, or injure soft tissues.


Fixation – lag screw as primary fixation device separate from plate enlarge

Tap the thread hole

Using a 3.5 mm cortical tap with the corresponding drill sleeve to protect adjacent soft tissues, create a thread in the pilot hole of the far cortex.
This maneuver is not necessary when using self-tapping screws.


Fixation – lag screw as primary fixation device separate from plate enlarge

Lag screw insertion

Insert the lag screw and carefully tighten it, making sure that the fracture stays reduced and is compressed.
Help to maintain the reduction by leaving the pointed reduction forceps in situ.


Fixation – lag screw as primary fixation device separate from plate enlarge

Neutralization (protection) plate

In order to protect the primary lag screw fixation, a perfectly contoured plate is applied adjacent to the screw head without axial compression (all screws in the neutral position).
The pointed reduction forceps can then be removed.

5 Fixation – lag screw as primary fixation device through the plate top

Fixation – lag screw as primary fixation device through the plate enlarge

In some fracture configurations, it might be advantageous to compress the fracture first by inserting a lag screw through the relevant plate hole.
If this technique is used, then axial compression must not be applied via the plate.


Fixation – lag screw as primary fixation device through the plate enlarge

Drill gliding hole through plate

After anatomical reduction, the fracture is held by a pointed reduction forceps.
A perfectly contoured plate is now applied to the surface of the anatomically reduced fracture and fixed to one main fragment, either with a neutral screw, or with a bone forceps as illustrated. The interfragmentary lag screw can then be inserted through the appropriate plate hole.
Using a 3.5 mm drill sleeve and a 3.5 mm drill bit, drill a gliding hole in the near cortex.
Ensure that the direction of the drill is as perpendicular to the fracture plane as possible.


Fixation – lag screw as primary fixation device through the plate enlarge

Thread hole: guide in gliding hole

Insert the sleeve of the 3.5 mm / 2.5 mm drill guide through the plate and into the gliding hole. Check that the anatomical reduction is maintained and use a 2.5 mm drill bit to drill a pilot hole just through the far cortex.


Fixation – lag screw as primary fixation device through the plate enlarge

Measure for screw length

Use a depth gauge through the plate to measure for screw length.
Measure the longer side of an oblique drill hole, as shown, to ensure sufficient screw length.
A screw should protrude 1-2 mm through the opposite cortex to ensure maximal thread purchase. However, too long a screw may be tender, or injure soft tissues.


Fixation – lag screw as primary fixation device through the plate enlarge

Tap the pilot hole

Using a 3.5 mm cortical tap with the corresponding drill sleeve to protect adjacent soft tissues, create a thread in the pilot hole of the far cortex.
This maneuver is not necessary when using self-tapping screws.


Fixation – lag screw as primary fixation device through the plate enlarge

Lag screw insertion

Insert the lag screw through the plate and carefully tighten it, making sure that the fracture stays reduced, is compressed, and that the plate sits accurately on the bone surface.


Fixation – lag screw as primary fixation device through the plate enlarge

Insert the plate screws

Complete the fixation by inserting the plate screws in neutral mode, usually alternately, in both main fragments.

6 Fixation – compression plating with an additional lag screw top

Fixation – compression plating with an additional lag screw enlarge

Prebending the plate

If the plate which will exert axial compression is exactly contoured to the anatomically reduced fracture surface, there will be some gapping of the opposite cortex, when the plate is tensioned by tightening the load screw.
The solution to this problem is to “over-bend” the plate so that its centre stands off 1-2 mm from the anatomically reduced fracture surface.
When the neutral side of the plate is applied to the bone, slight gapping of the cortex will occur directly underneath the plate.
As the load screw is tightened, the tension generated in the plate compresses the fracture evenly across the full diameter of the bone.


Fixation – compression plating with an additional lag screw enlarge

Prebend the plate

After the plate has been contoured anatomically to the reduced bone surface, prebend the plate with the handheld bending pliers, or a pair of bending irons.


Fixation – compression plating with an additional lag screw enlarge

Insert 1st screw

The prebent plate is fixed to one of the main fragments with a screw in neutral mode. A reduction forceps is placed on the opposite fragment to hold it in the reduced position against the plate.
The aim of this maneuver is to create an “axilla” into which the other fragment can be compressed and locked either with a compression screw or using a push-pull technique.

Note: Because of the design of the LC-DCP holes, the neutral drill guides for the LC-DCP have a very slightly eccentric hole and an arrow, which needs always to point towards the fracture line.


Fixation – compression plating with an additional lag screw enlarge

Maintain reduction

A reduction forceps is placed on the opposite fragment to hold it in the reduced position against the plate.


Fixation – compression plating with an additional lag screw enlarge

Option: Push-pull reduction technique
If reduction proves difficult, a bone spreader, placed between the end of a plate and an independent cortical screw, can be used to distract the fracture for reduction. Once reduced, the reduction forceps is tightened.


Fixation – compression plating with an additional lag screw enlarge

Then, using the same independent screw, axial compression can be obtained by pulling the plate end towards the screw with a small Verbrugge clamp.


Fixation – compression plating with an additional lag screw enlarge

Insert 2nd screw eccentrically

Once reduced, a second screw is inserted eccentrically (yellow drill sleeve) into the reduced fragment.

Note: the arrow on the drill sleeve has to point towards the fracture line.


Fixation – compression plating with an additional lag screw enlarge

Tighten screw

By tightening the eccentrically-inserted screw, axial compression is achieved.


Fixation – compression plating with an additional lag screw enlarge

Insert additional screws

Should it be necessary to increase the axial compression, a second screw can be inserted eccentrically into the same fragment as the first eccentric screw.
As the second eccentric screw is tightened, the first eccentric screw in the same fragment needs to be loosened slightly to allow the plate to slide on the bone.


Fixation – compression plating with an additional lag screw enlarge

Pearl: alternative drill sleeve

For inserting screws into the limited contact dynamic compression plate (LC-DCP), the Universal Drill Guide can be used as well. When this drill guide is pressed into the plate hole, the screw position will be neutral (A). When it is held against the end of the plate hole, without exerting downward pressure, the screw position will be eccentric (B).


Fixation – compression plating with an additional lag screw enlarge

Lag screw

The stability of a plate fixation in oblique fractures can be greatly increased by inserting a lag screw through the plate after primary axial compression has been achieved.


Fixation – compression plating with an additional lag screw enlarge

All other plate screws are now inserted in a neutral position (green drill sleeve) and do not serve further to increase compression.


Fixation – compression plating with an additional lag screw enlarge

Completed osteosynthesis

This illustration shows the completed compression plate osteosynthesis of a oblique fracture with additional lag screw fixation.

7 Check of osteosynthesis top

Check of osteosynthesis enlarge

Check the completed osteosynthesis by image intensification. Make sure that the plate is at a proper location, the screws are of appropriate length and a desired reduction was achieved.


Check of osteosynthesis enlarge

The elbow should be stabilized at the epicondyles and the forearm rotation should be checked between the radial and ulnar styloids.

8 Assessment of Distal Radioulnar Joint (DRUJ) top

Before starting the operation the uninjured side should be tested as a reference for the injured side.

After fixation, the distal radioulnar joint should be assessed for forearm rotation, as well as for stability. The forearm should be rotated completely to make certain there is no anatomical block.


enlarge

Method 1

The elbow is flexed 90° on the arm table and displacement in dorsal palmar direction is tested in a neutral rotation of the forearm with the wrist in neutral position.

This is repeated with the wrist in radial deviation, which stabilizes the DRUJ, if the ulnar collateral complex (TFCC) is not disrupted.


enlarge

This is repeated with the wrist in full supination and full pronation.


enlarge

Method 2

In order to test the stability of the distal radioulnar joint, the ulna is compressed against the radius...


enlarge

...while the forearm is passively put through full supination...


enlarge

...and pronation.

If there is a palpable “clunk”, then instability of the distal radioulnar joint should be considered. This would be an indication for internal fixation of an ulnar styloid fracture at its base. If the fracture is at the tip of the ulnar styloid consider TFCC stabilization.

v2.0 2013-07-09