Scapula 14.1 B1.1 Partial articular, anterior simple - lag screw fixation (MIO)

back to skeleton

Glossary

1 Principles/Introduction top

Main goal

In this procedure the main goal is to reconstruct the joint surface.


Screw selection enlarge

Screw selection

For this procedure the following screws are typically used:

  • Cannulated 3.5 lag screws
  • Headless 3.0 screws

Headless screws are preferable where fixation involves inserting them through capsule or labrum. 

If a headless screw is used, bicortical purchase is preferred but not required for compression.

2 Approach top

Approach enlarge

Anterior fragments are reduced and fixed through the minimal invasive anterior approach with the camera inserted posteriorly.

3 Reduction top

Reduction enlarge

Reduction of the articular surface may be facilitated by the insertion of a K-wire to be used as joystick. For this reason we prefer to use the cannulated system and insert the K-wire in such a way that it will subsequently serve as a guide for the lag screw trajectory.

The liquid used for distension should not be under high pressure. If it is under high pressure it will drive the fragments apart and make the reduction difficult.

If a satisfactory reduction cannot be achieved by closed manipulation one must resort to an open reduction and fixation.

Whenever K-wires are used as joysticks, whether as subsequent guide wires or not, they should not be inserted trans articulary.


enlarge

When reduction is completed, the joystick K-wire is now inserted to temporarily fix the fracture.

The pressure of the arthroscopy liquid can now be increased to normal if necessary.

Make sure that K-wires are not directed into the suprascapular notch where they can compromise the neurovascular bundle.

4 Fixation top

Fixation enlarge

An appropriate length lag screw is inserted.


enlarge

The K-wire is removed.

Depending on the size of the fragment, two lag screws are preferred since one does not offer rotational stability.

Correct reduction is continuously verified by the arthroscopic view.


enlarge

Check the position of the screws and the reduction by image intensification.

1.0