Malleoli 44-A1.3 Tension band wiring

back to skeleton

Glossary

1 Principles top

enlarge

The tension band converts tensile forces on the lateral side into compression forces.

In order to be able to use a tension band, the opposite cortex needs to have contact and be able to resist compression.


enlarge

The figure-of-eight wire loop lies on the lateral surface of the lateral malleolus and acts as a tension band when tightened.

Choose a wire of sufficient strength to withstand the tensile forces generated in the figure-of-eight loop (0.8 mm).

2 Reduction top

enlarge

Reduction with small pointed reduction forceps

Reduce the fracture with small pointed reduction forceps.


enlarge

Preliminary stabilization

Drill a 1.6 mm K-wire parallel to the fibular axis into the anterior portion of the tip of the lateral malleolus using a 2.0 mm drill sleeve.

Insert the K-wire until its point reaches a minimum of 2.0 cm proximal to the antero-posterior drill hole for the wire.


enlarge

Insert a second K-wire

A second 1.6 mm K-wire is inserted parallel to the first one into the posterior portion of the tip of the lateral malleolus.

These two parallel K-wires stabilize the fracture against rotation and will anchor the figure-of-eight wire distally.

Check the reduction under image intensification.

Note

Care must be taken to avoid intraarticular insertion of the K-wires - check specifically with the image intensifier.

3 Fixation top

enlarge

Drilling

Drill an antero-posterior hole with the 2.0 mm drill bit and protection sleeve, transverse to the long fibular axis, approximately 3 cm proximal to the fracture.

Keep soft-tissue dissection to a minimum.


enlarge

Pearl

In osteoporotic bone, in order to avoid cutting out of the wire from the drill hole, a bicortical screw may be used as an alternative means of anchoring the proximal end of the wire.


enlarge

Wire preparation and insertion

Prepare the 0.8 mm wire by making a loop approximately one third along its length.

Insert the shorter segment of the wire through this drill hole.


enlarge

Figure-of-eight configuration

The small pointed reduction forceps can now be removed.

Pass the long segment of the wire (bearing the loop) in a figure-of-eight configuration around the protruding ends of the K-wires.

Twist the free ends of the wire firmly together.


enlarge

Pearl

Some surgeons prefer to use two pieces of wire, twisting the ends together to create the figure of eight loop with a twist on each side.


enlarge

Tightening the wire

The wire twist is loosely prepared ensuring that each end of the wire spirals equally, forming a double helix – avoid twisting one wire around the other (straight) wire.


enlarge

The wire is tensioned by pulling on the twists.


enlarge

The slack is then taken up by further twisting. Repeat this until the desired tension is achieved. Both loops must be tightened at the same time and in the same direction, in order to achieve equal tension on both arms of the wire.

Note

In osteopenic bone care must be taken to avoid excessive tensioning.


enlarge

Prevent later soft-tissue irritation

Trim the twists and turn the ends towards the fibula in order not to irritate the soft-tissues later.


enlarge

Sinking the K-wires

Cut the 2 K-wires obliquely approximately 1 cm from their insertion points and, with the help of a bending iron and forceps, bend through 180 degrees.
The K-wires are then driven home, sinking their curved ends into the bone in order to prevent backing out and skin injury.

Take final x-rays in both planes to check results.