Executive Editor: Steve Krikler

Authors: Paulo Barbosa, Felix Bonnaire, Kodi Kojima

Malleoli 44-B1.3 Antiglide plate

back to skeleton

Glossary

1 Principles top

enlarge

The one-third tubular plate, which is placed posterolaterally on the fibula in an antiglide position, indirectly reduces the fracture and acts as a buttress to resist the posterior and proximal displacement of the distal fragment.

Additionally, the main fracture can be compressed with a lag screw, inserted through the plate.

The third fragment, if large enough, can be fixed to the main lateral malleolar fragment with a cancellous lag screw. This technique is used for the three part type of 44-B1.3 fracture, in which an oblique fracture of the fibula has occurred through the transsyndesmotic zone, and the tip of the proximal fragment ("ski tip") has fractured off.

2 Preliminary reduction top

enlarge

Preparation of the fracture site

Check that the third fragment has not become trapped between the main fragments. This can be a cause of irreducibility of an apparent 44-B1.2 fracture. The other clue is often anterior tibio-fibular ligament fibres disappearing into the main fracture site.

Do not strip the periosteum more than necessary.

 


enlarge

Preliminary reduction before plate application

Having released the third fragment from between the main fibular fragments as necessary, reduce the fracture by applying light longitudinal traction and rotating the foot internally.

3 Plate preparation top

enlarge

Bending the plate

Prepare a four hole one-third tubular plate. In order to improve the anti-glide buttressing effect, the plate is slightly bent at its distal end.

This applies posterior compression on the distal fragment, which augments compression across the fracture plane.


enlarge

Planning the plate position

Accurate positioning of the plate is essential in obtaining the desired buttress effect and allowing the insertion of a lag screw. The proximal hole closest to the fracture site should be located 3-5 mm proximal to the fracture line. Ensure that the lower end of the plate is not too distal, as, without the integrity of the "ski tip" of the proximal fibular fragment, a too distal point of pressure on the malleolar fragment risks tilting it anteriorly.

4 Fixation top

enlarge

Drilling

Reduce the main fragments and hold with a pointed reduction forceps. Ensure that the third fragment is reduced before tightening the forceps.


enlarge

Either make sure the position of the forceps allow the plate to lie in the correct position, or alternatively, place the forceps through the hole in the plate before tightening the forceps.


enlarge

In the proximal fibular fragment, drill a 2.5 mm hole through both cortices, from posterior to anterior, 3-5 mm proximal to the apex of the fracture.


enlarge

Plate application

Carefully position the plate so that two holes overlie each fragment, and hold it in place with a finger.


enlarge

Insertion of the first proximal screw

Measure the length of the drilled hole through the plate.

Tap with the 3.5 mm cortical tap and insert the first 3.5 mm cortex screw.

Tightening the screw will push the plate against the fracture, holding it reduced.

Check reduction visually and under image intensification.


enlarge

Insertion of the second proximal screw

Insert the second screw through the most proximal plate hole, drilling, measuring and tapping through the plate hole.


enlarge

Prepare insertion of a lag screw

Drill a gliding hole through the plate hole closest and just distal to the fracture, using a 3.5 mm drill bit. The hole should be drilled as perpendicularly to the fracture line as possible.

Insert a 2.5 mm drill sleeve inside the gliding hole and drill the threaded hole in the far cortex with a 2.5 mm drill bit.

Measure the depth of the hole with the hook of the depth gauge pointing proximally.

Tap the far cortex with the 3.5 mm cortical tap.


enlarge

Insertion of the lag screw

Insert a 3.5 mm cortical screw, functioning as a lag screw, and tighten carefully.

If the bone is osteoporotic, or if the fragment is large, a second cortical lag screw is inserted through the remaining distal plate hole.

Final x-rays should be taken now.


Note

In less oblique fractures, or with a large third fragment, it is not possible to insert a lag screw. In this case you can omit the postero-anterior lag screw.


enlarge

Fixation of third fragment

The final stage is the fixation of the third fragment to the main lateral malleolar fragment, using an anteroposterior cancellous lag screw, fitted with a washer.

v1.0 2006-12-04