Executive Editor: Ernst Raaymakers, Joseph Schatzker, Rick Buckley

Authors: Matthias Hansen, Rodrigo Pesantez

Proximal tibia 41-A2 Direct reduction

back to skeleton

Glossary

1 Principles top

enlarge

Absolute stability

In case of a simple fracture pattern absolute stability is mandatory. You can achieve this with lag screw technique. Depending on the fracture pattern the lag screw could be through the plate (illustrated here) or outside the plate.


enlarge

For such fracture pattern where the lag screw is outside the plate, the lag screw is applied first then a plate is used as a neutralization plate.

2 Reduction top

enlarge

Clamps

Indirect reduction may be achieved by external manipulation of the fractured fragment using clamps. In cases where adequate closed reduction is not achieved, open anatomical reduction is required.


enlarge

Pearl: clamp on plate

If you intend to keep the clamp on throughout the whole procedure it is best to slip the plate under the clamp prior to tightening the clamp to maintain reduction. Under these circumstances you determine which screw hole is best for the placement of the tip of the clamp from the pre-operative plan and intra-operative trial.

3 Fixation top

enlarge

Lag screw through the plate

First, the lag screw is applied through the plate using the conventional lag screw technique.

Click here for a detailed description of the lag screw technique.


enlarge

Additional screw insertion

If possible three bicortical screws on each side of the fracture should be applied. No particular order of insertion has to be followed.


enlarge

Alternative: lag screw outside the plate

Illustration showing the lateral projection of the use of a lag screw outside the plate.

v2.0 2010-05-15