Mounting the nail on the insertion handle enlarge

Mount nail on insertion handle

The humeral nail is mounted on an insertion handle. The nail must be rotated correctly relative to the insertion handle.

If an angled nail is used (as illustrated) ensure that the apex of the nail curvature points away form the insertion handle.


Nail insertion enlarge

Insert nail and reduce fracture

Insert the nail with slightly rotating movements down to the metaphyseal fracture line. Pass the fracture zone under image intensification and make sure that the nail enters the distal fragment properly.


Correct placement of nail enlarge

Make sure the proximal end of the nail is placed beneath the bony surface of the humeral head.
No protrusion of the nail may be tolerated. Confirm with appropriately oriented C-arm images that the nail is below the bone.
The nail illustrated has a transverse spiral blade. At the recommended depth, the spiral blade should lie between the middle and lower third of the humeral head. Depending upon the selected humeral nail, different preoperative planning is necessary for its locking devices.
Be careful not to displace tuberosity fractures when inserting proximal locking components.


Retrotorsion of the humeral head enlarge

Retrotorsion of spiral blade/screw

In order to lock the spiral blade, mount the aiming arm and swivel it approximately 25° anteriorly in order to follow the retroverted axis of the humeral head. (The humeral head axis is directed approximately 25° posteriorly to the condylar plane of the distal humerus.)


Spiral blade position enlarge

Within the humeral head, the spiral blade should be placed at the transition of the middle to the lower third, slightly below the equator.


Mounting aiming device and insertion of trocar combination enlarge

Mount aiming device and insert trocar combination

Mount the aiming device in the insertion handle. Confirm that the retroversion angle is correct. Make a skin incision for the aiming device, dissect the muscles bluntly down to the bone, and fully insert the trocar.


Guide wire insertion enlarge

Insert guide wire

Check once more the retrotorsion of the handle.
Remove the central trocar and drill the guide wire for the spiral blade onto the medial cortex of the humeral head.
The position of the guide wire can be checked under image intensification.
Be aware not to perforate the humeral head in order not to insert a too long spiral blade.


Determination of correct spiral blade length enlarge

Determine length of spiral blade

Determine the correct length of the spiral blade with the appropriate depth gauge.


Opening lateral cortex enlarge

Open the lateral cortex

Perforate the lateral cortex with the appropriate cannulated drill.


Spiral blade insertion enlarge

Insert spiral blade

Attach the spiral blade to the inserter and introduce both over the guide wire.
Align the handle of the inserter parallel to the aiming arm.
The initial rotation of the T-handle of the spiral blade inserter relative to the aiming arm depends on patient anatomy. If the distance from the lateral cortex to the nail is less than 10 mm, start the inserter slightly clockwise from parallel. If the distance from the lateral cortex to the nail is more than 10 mm, start the T-handle slightly counter-clockwise from parallel.
By applying light controlled hammer blows to the connecting screw, advance the spiral blade to the desired depth. This causes the handle to rotate 90°.


tension band sutures for the rotator cuff enlarge

Pearl: tension band sutures for the rotator cuff
Place these in the base plate of the spiral blade before the blade is fully inserted.

Monitor the depth of the spiral blade with image intensification. If attaching sutures to the spiral blade, pause when the spiral blade is approximately 1.5 cm to 2.0 cm short of its intended position so that the suture ends can be placed through the appropriate holes in the base blade.


Checking position of spiral blade enlarge

Check position of spiral blade

Check the position of the spiral blade by image intensification.

Pitfall: captured bicipital tendon
If the chosen nail has locking screws that might pass through the bicipital groove, be careful that they do not trap the bicipital tendon.


Locking screw enlarge

Drill and determine length of locking screw

For distal locking, insert the two-piece trocar combination (aiming arm). Through an appropriately placed trocar, drill through both humeral cortices until the bit just breaks through the medial cortex, and read the depth from the drill bit. Alternatively, a depth gauge can be used.
Insert a locking screw through the trocar. A second screw is recommended, especially in osteoporotic bone.

Pearl: make one incision large enough to allow palpation of the axillary nerve.


End cap insertion enlarge

Insert the end cap

The end cap prevents tissue from plugging the inner thread of the nail. Furthermore it offers the option of angular stability, by compressing the spiral blade.
End caps are available in different sizes and can, if necessary, be used to extend the nail. The top of the end cap must not protrude above the surface of the bone.


Supplementary rotator cuff tendon sutures enlarge

Supplementary rotator cuff tendon sutures

Rotator cuff tension band sutures are placed using the sutures previously placed in the spiral blade base plate.


Repair of rotator cuff enlarge

Repair rotator cuff

Suture the supraspinatus split.