enlarge

Plate application

Drill a 2.5 mm hole through both fibular cortices at the planned screw site.

Measure the screw length through the plate, and tap with the 3.5 mm cortical tap in its tap sleeve.

Carefully apply the plate. Insert the first proximal screw. The screw should just penetrate the far cortex.


enlarge

Insertion of the first distal screw

Next, insert a screw into the first distal plate hole. All distal screws are aimed towards the articular surface of the lateral malleolus. Ensure that their tips do not protrude into the joint by aiming the drill slightly posteriorly.

Carefully predrill an eccentric hole (at the distal edge of the plate hole) until the drill bit just penetrates the far cortex. Measure the length and select a cortex screw approximately 2 mm shorter. Tap threads only into the near fibular cortex and then insert the screw.

Note
In osteopenic bone, fully threaded cancellous bone screws can be used in the distal fragment.


enlarge

Completing screw insertion

Insert the remaining screws as described above in the sequence of the preoperative plan.

The most distal screw is inserted slightly obliquely, from distal to proximal, in a posterior direction. Again, choose the screw 2 mm shorter than measured and tap the threads only in the near cortical bone.

Gently tighten all screws.

Check under image intensification in both planes to ensure that no screw protrudes into the ankle joint or the syndesmosis.


enlarge

Pearl

A one-third tubular plate molded as a hook plate can be used to fix a small lateral fragment (where there is space for only one screw), or a severely osteoporotic bone.

The sequence of the application of the hook plate differs from the procedure described above.

First, place the tip of the hook through the ligament into the tip of the lateral malleolus.

Use an impactor for anchoring the hook tip in the bone.

Insert a screw into the hole immediately above the hook to secure the plate onto the distal fragment.

Insert a proximal cortical screw in an eccentric position to obtain compression.

Complete the fixation by inserting the other screws in the planned sequence.